Sirona Dental Systems GmbH
6 Installation
Installation Instructions Facescan
6.7 Attaching the scan unit
båÖäáëÜ
5. Turn the holding plates (G) over the chassis plate of the face scan
unit.
6. Push the installation aid (C) with the scan unit on to the sensor
surface (H) of the X-ray detector.
7. Align the installation aid (C) to the center of the sensor surface (H).
8. Move the safety screw (E) on both sides from the parked position (L)
(delivered condition) to the function position (M).
9. Screw the scan unit down from below with the safety screws (E). In
so doing, the screws should pierce through the housing of the X-ray
detector.
The distance (K) between the plate of the scan unit and the cover
of the X-ray detector should be 6 mm.
10. Screw down the 6 screws (D).
IMPORTANT
Possible faulty alignment
Displacement of the installation aid (C) can be caused through further
fastening of the scan unit with the safety screws (E).
➢ When screwing in the safety screws (E), please ensure that the
installation aid (C) always remains aligned.