Sirona Dental Systems GmbH
9 Appendix
Installation Instructions Facescan
9.2 Conversion instructions for non-"Facescan ready" units
båÖäáëÜ
9.2.6
Attaching board DX32
➢ Attach board DX32 in reverse sequence to that described in section
"Removing board DX32 [ → 54]".
9.2.7
Connecting switch S1
Disconnecting cable L2 from switch S1
1. Remove cable L2.BU from switch S1.4.
2. Remove cable L2.BN from switch S1.1.
Connecting cable L2 to switch S1
1. Insert cable L2.BU into the flat power strip of cable L77(L70.GN)
(green).
2. Insert cable L2.BN into the flat power strip of cable L77(L70.BK)
(black).
IMPORTANT
Add cable ties
For the following steps, all necessary cable ties that were removed
previously are to be replaced.
IMPORTANT
Assembly instructions
Before assembling board DX 32, the bracket (C) needs to be
reattached.