7 Commissioning
Sirona Dental Systems GmbH
7.2 Connection to the network
Installation Instructions Facescan
Checking the configuration
1. Insert the FaceScan USB stick into a PC.
2. With a text editor program, open the log file "facescan_settings.log".
3. Check the entries in the log file.
If the configuration has been successful, the log file should state:
"Network configuration successful!"
4. Remove the USB stick from the PC.
Closing the FACESCAN USB slot
Closing FaceScan
1. Pull cable L78(L74) from the X2 socket of the FACESCAN modular
board.
2. Screw the protective plate (T) onto the FaceScan unit.
3. Plug cable L78(L74) into the X2 socket on the FACESCAN modular
board.
4. Screw down the cover (A) onto the FaceScan.