63 76 482 D 3572
D 3572
.
031.01.04
.
02 12.2015
45
Sirona Dental Systems GmbH
7 Commissioning
Installation Instructions Facescan
7.6 Test exposure
båÖäáëÜ
7.6.3
Example 3
The exposure is not okay.
● Artifact (E) (very wavy surface.)
Reason: Too light or lighting too direct.
● Artifact (F) (strong duplication effect of the structure (see Example
1 [ → 43], artifact (B)).
Reason: Device is not calibrated or poor lighting conditions during
exposure.