Installation Instructions Facescan
Connecting power supply cable L77(L70) (to terminal K1) with switch
S1
1. Connect power supply cable L77(L70.BU) (blue) to switch S1 above
left.
2. Connect power supply cable L77(L70.BN) (brown) to switch S1
above right.
3. Connect power supply cable L77(L70.GN) (green) with flat power
strip to switch S1 below left.
4. Connect power supply cable L77(L70.BK) (black) with flat power strip
to switch S1 below right.
5. Connect power supply cable L77(L70.PE) (green/yellow) to the
ground point (E).
IMPORTANT
Keep safety clearance!
Both flat power strips (A) of power supply cable L77(L70) must be
attached to the lower connections of switch S1. Here you must ensure
that the open strip contacts (B) point downwards.