6 Installation
Sirona Dental Systems GmbH
6.4 Installing PoE module
Installation Instructions Facescan
6.4
Installing PoE module
Preparation
CABA-No. 7000000086
1. DANGER! Potentially lethal shock hazard! Disconnect the
GALILEOS from the junction box of the building installation.
2. Only applicable for units prepared in the factory: Cut through the
cable tie (A) on the Ethernet cable L77(L72).
Securing
1. Insert two plastic clamping terminals into the shaft (S).
2. Tip: Prevent the plastic clamping terminals from slipping using
adhesive tape.
3. Put a serrated washer on the thread bolts (H).
4. Put the PoE module (P) on the thread bolts (H) of the mains filter plate
(J) (the serrated washer must sit behind the fitting plate of the PoE
module).
5. Screw down the PoE module (P) to the place provided in the stand
with two screws (K) and a nut (L).
IMPORTANT
Possible assembly errors
➢ Make sure that cable L76 is behind the mains filter plate (J).