Sirona Dental Systems GmbH
9 Appendix
Installation Instructions Facescan
9.2 Conversion instructions for non-"Facescan ready" units
båÖäáëÜ
9.2.8
Removing the laser module
1. Remove the plate (A) with the laser module.
2. Allow the plate (A) to hang carefully
9.2.9
Removing the cable conduit with cable L13 on the ring
1. CAUTION! Risk of injury! The covering plate can have sharp edges.
Remove the covering plate (A) carefully from the X-ray detector
upwards from below.
2. Remove cable L13 from connector X201 (F) on board DX89.
3. Pull cable L13 from socket X500 on board DX1.
4. Remove cable L13 along with the cable conduit from the ring (see the
GALILEOS Service Manual)