7 Commissioning
Installation Instructions Facescan
7.4 Calibration
båÖäáëÜ
Configuration
✔
The
"Facescan configuration"
dialog box opens.
1. Press the
"New device"
button.
A password dialog box appears.
2. Enter "service" in the user field.
3. Enter "sirona" in the password field.
4. Confirm the password dialog.
5. In the
"Name:"
text field, name the respective FaceScan unit.
6. In the
"HTTP address:"
text field, enter the IP address of the
FaceScan.
7. In the
"MAC address:"
text field, enter the MAC address of the
FaceScan.
8. In the
"GALILEOS:"
selection field, select the GALILEOS unit on
which the FaceScan is installed.
9. Confirm your settings by pressing the
"OK"
button.
7.4
Calibration
Explanation
Essentially, once the Facescan has been installed, a re-calibration of the
GALILEOS unit has to be performed.
The Facescan is automatically calibrated during the 3D calibration of
GALILEOS using the 3D calibration phantom (A).
Prerequisites
● The FaceScan has been correctly assigned to the GALILEOS (via
FaceScan configuration).
Calibration
The calibration is performed with the geometric calibration of GALILEOS
(see the GALILEOS Installation Instructions).
A