63 76 482 D 3572
D 3572
.
031.01.04
.
02 12.2015
37
Sirona Dental Systems GmbH
7 Commissioning
Installation Instructions Facescan
7.2 Connection to the network
båÖäáëÜ
7.2.1.3.1
Syntax of the configuration file "facescan_settings.cfg"
Configuration without DHCP
Text entry (factory setting):
Configuration with DHCP
Text entry:
7.2.1.4
Type 2: Configuration over a network cable (peer-to-peer)
Connecting the Facescan unit to a PC
✔
A PC with an installed web browser must be available.
✔
The factory setting of the IP address of the Facescan is
192.168.16.240.
✔
The PC employed must be in the 192.168.16.xx network; otherwise,
no network connection can be achieved.
➢ Using a network cable, connect the PC directly to the GALILEOS
media converter with installed Facescan.
Opening the web dialog / general types 1
Opening the web dialog
1. Switch GALILEOS on.
2. Open up a web browser on the PC.
3. Enter the Facescan IP address into the web browser
(http://<IP-Adresse>).
A password dialog box opens.
4. In the field
"User"
enter
"service"
.
5. In the field
"Password"
enter
"sirona"
.
The
"Facescan Device Service "
web dialog opens.
IMPORTANT
Pay attention to the syntax!
Text entries should never have a space before and after "=".
Examples:
Correct: USB_STICK_MODE=CONFIG
Incorrect: USB_STICK_MODE= CONFIG
IMPORTANT
Changes to network addresses.
➢ Adjust the entries
"IP"
(IP address) and
"Netmask"
(subnet mask)
as required.