9 Appendix
Sirona Dental Systems GmbH
9.2 Conversion instructions for non-"Facescan ready" units
Installation Instructions Facescan
9.2.2
Removing board DX32
1. Unscrew the bracket (C). The bracket (C) is located in front of board
DX32.
2. Move the stand to a height of 1260 (control panel display)
3. DANGER! Potentially lethal shock hazard! Disconnect the
GALILEOS from the junction box of the building installation.
4. Remove the cable ties (A) from cable L2.
5. Loosen all the left-hand screws (B) of the protective plates (E).