9.1
Servicing schedule
WHEN
WHO
WHAT
PURPOSE
Every 3
months and at
every oppor-
tunity
Operator
Leakage check on the
measuring cell KPFL 30.
Section 9.2
Preventing corrosion and main-
taining measuring accuracy.
Every 3 mon-
ths
Operator
Recalibrate the photome-
ter. Section 9.3
Measure for maintaining meas-
uring accuracy. Interval depen-
dent on medium.
Every 6
months or as
needed
Operator
Check and any cleaning of
the measuring cell KPFL
30. Section 9.4
Obligatory measure for main-
taining measuring accuracy.
Every 2 years
or as needed
Service
center
Replace the seals on the
closed measuring cell. Sec-
tion 9.5
Obligatory measure for main-
taining functional efficiency.
Every 5 years
or as needed
Operator /
service
center
Replace the measuring cell
windows on the closed
measuring cell KPFL 30.
Section 9.5
Measure for maintaining meas-
uring accuracy. Interval de-
pendent on medium and oper-
ating conditions.
Every 6
months or as
needed
Operator
Replace the pump mem-
brane according to the
Service & Operating Man-
ual VERSA-MATIC.
Section 1.4
Obligatory measure for main-
taining functional efficiency.
Annually or as
needed
Operator
Clean the free-fall measur-
ing cell KPFLJC PVDF.
Section 9.6
Obligatory measure for main-
taining measuring accuracy. In-
terval dependent on medium.
Every 2 years
Operator
Replace the light source.
Section 9.8
Obligatory measure for main-
taining measuring accuracy.
Every 5 years
Operator
Replace the fan.
Section 9.9
Maintaining the cooling in the
housing.
As needed
Operator
Replace the inlet pipes:
Inlet pipe KPFLJC PVDF
Section 9.10 / Inlet pipe
KPFLJ VA Section 9.11 /
Short inlet pipe PVDF of
sample conditioning sys-
tem Section 9.12
Obligatory measure for main-
taining functional efficiency.
As needed
Operator
Clean the sample prepara-
tion system and replace
the cover seal.
Section 9.13
Obligatory measure for main-
taining functional efficiency.
As needed
Operator
Clean the pipes.
Section 9.14
Obligatory measure for main-
taining functional efficiency.
Table 1: Servicing schedule