Instruction Manual OilGuard 2 Ex
Mounting
28
13003E/3
For the montage of the sample feed pump according to the following process:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Preassemble the pump bracket on the stand
using the four screws.
2.
Position the pump on the top pump bracket.
3.
Adjust the position of the pump by moving the
pump bracket.
Fix the pump bracket in place by tightening the
screws (circles).
4.
Mount the pump inlet pipe between the pump
outlet (Figure 4, pos. 8) and the inlet on the
sample preparation system.
5.
Connect the sample feed to the pump inlet
(Figure 4. pos. 2) (provided by the customer).
The pump inlet nozzle is provided with an in-
ternal thread 1/2 "NPT (arrow).
6.
Attach the earth (Figure 4, pos. 6) to the
pump.
The earth is fixed to the stand on one side and
on the foot of the pump on the other.