Mounting
Instruction Manual OilGuard 2 Ex
13003E/3
27
4.5
Mounting the sample feed pump (optional)
EXPLOSION
HAZARD!
Explosion hazard due to static discharge of the pumps.
Static discharges can occur when the pumps are in operation. This can lead to explosions.
In potentially explosive areas, the pumps must be connected to earth.
A sample preparation system must always be in place when a sample feed pump is used.
This is because the sample preparation system is used as a pulsation damper. Section 2.3
Figure 4: Sample feed pump
Sample feed pump
Pump inlet
Needle valve for fine adjustment of
the pump cadence
Pressure display
Pressure reducing valve
Earth
Ball valve
Pump outlet