Servicing
Instruction Manual OilGuard 2 Ex
13003E/3
67
9.3
Recalibrating the photometer
9.3.1
Preparing for recalibration using the checking unit
The instrument must be at operating temperature for recalibration. The warm-up time lasts
2 hours.
Figure 19: Checking unit on OilGuard 2
Fluorescence glass
Locking screws
Serial number of the photometer
Nominal value
The following procedure describes how to prepare the instrument for recalibration using the
checking unit.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample feed and remove the inlet.
2.
Remove the cover of the measuring cell.
On the closed measuring cell KPFL 30, con-
tinue with step 5.
On the free-fall measuring cell
KPFLJC PVDF, carry out steps 3a and 4.
On the free-fall measuring cell KPFLJ VA,
only carry out step 3b.