Servicing
Instruction Manual OilGuard 2 Ex
13003E/3
75
9.6
Contamination check on the free-fall measuring cell
KPFLJC PVDF
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt both the sample feed and return, re-
move them from the measuring cell and let the
cell run dry.
2.
Loosen the locking screws (arrows) and remove
the cover on the measuring cell.
3.
Remove the lock nut (B) by turning counter-
clockwise (bayonet lock) and pull out the inlet
pipe (A).
This measuring cell is available with a
long or short inlet pipe. This example uses a
long inlet pipe. There is no difference in han-
dling here.
4.
Pull out the protective pipe (A).