Mounting
Instruction Manual OilGuard 2 Ex
13003E/3
31
4.7
Mounting the inlet pipe on the free-fall measuring cell
PVDF
Longer inlet pipes are used on measuring systems with no sample preparation system. Sec-
tion 2.3
The following procedure describes the installation of the supplied inlet pipe. In the example
provided, a long inlet pipe is used. The procedure is the same if a short inlet pipe is used.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Insert the protective pipe (A) from above into
the measuring cell housing.
The pin on the protective pipe (B) must be in-
serted in the slot on the measuring cell
housing (C).
2.
Insert the inlet pipe (arrow) into the protective
pipe.
3.
Guide the lock nut over the inlet pipe and fix it
into place with a small turn clockwise (bayo-
netlock).