Troubleshooting
Instruction Manual OilGuard 2 Ex
13003E/3
93
10.1.2
Fault messages and effect on operation
FAULT
If a fault occurs during operation, it has the following
effects:
The measurement is canceled.
The measuring values go to 0. When the cause of
the fault has been remedied, it is automatically de-
leted.
When the
Fault
message appears,
the color of the
status display changes to
red
and the text de-
scribes the fault in question.
Example:
FAULT SERIAL 1
The following fault messages can be displayed:
FAULT MESSAGE
DESCRIPTION
POSSIBLE CAUSES
SERIAL 1
No communication between
the operating electronics and
measuring electronics.
Cable disconnected.
Defect in the electronic system.
Service technician
MEASUR.FAULT
Measuring value acquisition is
faulty.
Bubbles in the water.
External light in the vicinity of
the measuring point (e.g.
transparent hoses).
Defect in the electronic system.
Service technician
LIGHTSOURCE 1
The detector for monitoring
the light source receives no
light.
Defective light source.
Obstacle in the optical path.
Service technician
IO PORT
Connection between
NG_HAUPT and NG_BASI-
Print is disturbed.
Connection has been dis-
connected.
Defect in the electronic system.
Service technician
Table 4: Possible fault messages