Servicing
Instruction Manual OilGuard 2 Ex
13003E/3
85
9.12
Replacing the short inlet pipe PVDF (sample conditioning
system)
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample feed.
2.
Remove the connecting piece (C) between the
measuring cell and sample conditioning sys-
tem (A) by loosening the nuts (B and F).
3.
Remove the lock nut (E) by turning counter-
clockwise (bayonet lock) and pull out the short
inlet pipe PVDF (D) from the measuring cell.
4.
Insert the new short inlet pipe PVDF (D) and fix
in place with the lock nut (E).
5.
Insert the connecting piece (C) between the
measuring cell and sample conditioning sys-
tem (A) and fasten in place with the nuts
(B and F).