Instruction Manual OilGuard 2 Ex
Servicing
84
13003E/3
9.11
Replacing the inlet pipe on the KPFLJ VA
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample feed.
2.
Loosen the locking screws (circles) and remove
the old inlet pipe (A).
3.
With the short end first, insert the new inlet
pipe (A) into the measuring cell in the direction
of the arrow, then fasten in place with the
locking screws.
4.
Reattach the sample feed.