Instruction Manual OilGuard 2 Ex
Servicing
72
13003E/3
9.5
Replacing the measuring cell windows and seals on the
measuring cell KPFL 30
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove both the sample feed and return from
the measuring cell and let the cell run dry.
2.
Loosen the locking screws (arrows) and remove
the measuring cell by tipping it slightly.
3.
Loosen the Allen bolt (circle) and open the
clamp.
4.
Remove the measuring cell cover (A).
Содержание OilGuard 2 Ex