Instruction Manual OilGuard 2 Ex
Servicing
66
13003E/3
9.2
Leakage check on the measuring cell KPFL 30
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Carry out a visual check of the sample lines
(connecting pieces, connections etc.).
2.
Check for discoloration of the desiccant in the
inspection glass (arrow).
If the desiccant is colored orange, then
the desiccant cartridge has to be replaced or
regenerated.
Ruby red = dry / orange = damp
3.
Enter the inspection in the servicing log.
Section 16.1