Instruction Manual AquaMaster
Servicing
66
12745E/4
8
Servicing
8.1
Servicing schedule for AquaMaster
WHEN
WHO
WHAT
PURPOSE
Every three
months or as
needed
Operator
Cleaning, checking and, if
required, recalibration of
the pH sensor.
Section 8.1.4
Obligatory measure for main-
taining measuring accuracy.
Every three
months or as
needed
Operator
Cleaning, checking and, if
required, recalibration of
the conductivity sensor.
Section 8.1.5
Obligatory measure for main-
taining measuring accuracy.
Every three
months or as
needed
Operator
Cleaning, checking and, if
required, recalibration of
the redox/ORP sensor.
Section 8.1.6
Obligatory measure for main-
taining measuring accuracy.
Every three
months or as
needed
Operator
Cleaning, checking and, if
required, recalibration of
the oxygen sensor.
Section 8.1.7
Obligatory measure for main-
taining measuring accuracy.
As needed
Operator
Replacing sensors.
Replacing a sensor config-
ured by SIGRIST.
Section 8.1.8
Replacing an unconfigured
sensor.
Section 8.1.9
Obligatory measure for main-
taining measuring accuracy.
As needed
Operator
Cleaning the pipes and
parts which come into
contact with water. Sec-
tion 8.1.11
Measure for maintaining meas-
uring accuracy. Interval de-
pendent on water quality and
handling.
As needed
Operator
Cleaning the measuring
cell block.
Section 8.1.10
Measure for maintaining meas-
uring accuracy. Interval de-
pendent on water quality and
handling.
Table 1: Servicing schedule