Mounting and installation
Instruction Manual AquaMaster
12745E/4
33
WORKSTEP
ADDITIONAL INFO / IMAGES
2.2: Connect the cable as follows to the termi-
nals of the AQ2Basi print.
Pos. 1) Connection to the connection box
Terminals
8
9
16
17
Name
24V
GND
A
B
Cable color
Brown
Green
White
Yellow
Pos. 2) Connection to the Powerbox
Terminals
10
11
12
13
Name
SDA
GND
SCL
GND
Cable color
Gray
Pink
Blue
Red
Pos. 3) 0/4 .. 20 mA input
Terminals
24 (-)
25 (+)
26 (-)
27 (+)
Cable color
Black
Violet
Gray/
pink
Red/
blue
3.
Connect the
standard current outputs
ac-
cording to the following table:
AQ2Basi print:
Terminals
18
19
20
21
Name
mA 1 -
mA 1 + mA 2 -
mA 2 +
I/O module:
Terminals
28
29
30
31
Name
mA 3 -
mA 3 + mA 4 -
mA 4 +
4.
Connect the
optional current outputs (4-
way module)
according to the following ta-
ble:
4-way module:
Terminals
1
2
3
4
Name
mA 5 -
mA 5 + mA 6 -
mA 6 +
Terminals
5
6
7
8
Name
mA 7 -
mA 7 + mA 8 -
mA 8 +
5.
Connect the
two relay outputs
according to
the following table:
AQ2Basi print:
Terminals
1
2
3
4
5
6
Function
I/O
I/O
C
I/O
I/O
C
Name
Rel. 1
Rel. 2