Instruction Manual AquaMaster
Servicing
88
12745E/4
The optics unit is mounted on the docking station as follows:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Open the mounting clips (circle in figure be-
low) as follows:
Use a little force to push the red fuse in the
direction of the arrow (picture 1) and at the
same time lift the mounting clip (picture 2).
Press the mounting clip in the direction of the
arrow over the lock plate of the optics unit
(picture 3) and then open (picture 4).
Position of the mounting clips
Picture 1
Picture 2
Picture 3
Picture 4
2.
Remove the optics unit (Figure 19, pos. 1)
from the photometer, position it on the dock-
ing station (Figure 19, pos. 2) and then fasten
in place with the knurled screw.