Servicing
Instruction Manual AquaMaster
12745E/4
91
8.2.3
Flow rate reduction
A reduction of the minimum flow rate from 2.5 l/min to 1.3 l/min is possible; however, note
the following points:
Reproducibility worsens (from ±1% at 2.5 l/min to ±3% at 1.3 l/min).
The instrument must be very precisely leveled. If there is a 1° angle, the measuring value
can change by up to ±3%.
If the flow rate is low, it is even more important to maintain a constant supply flow be-
cause the fluctuations have a greater impact on the measuring value.