Servicing
Instruction Manual AquaMaster
12745E/4
103
8.2.8
Replacing the air filter
The following describes the replacement of the air filter:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the sample supply to the photome-
ter.
Section 5
2.
Interrupt the service voltage to the photome-
ter.
Section 4.4
3.
Remove the optics unit of the photometer
and fasten onto the docking station.
Section 8.2.1
4.
Remove the screws (circles) used for securing
the air filter and remove the filter cover and
filter (X).
5.
Replace the old filter with a new one and fas-
ten it onto the photometer again along with
the filter cover.
6.
Place the optics unit back on the measuring
cell unit and lock with the four mounting
clips.
Pay attention to the guide pins (see picture).