Servicing
Instruction Manual AquaMaster
12745E/4
75
WORKSTEP
ADDITIONAL INFO / IMAGES
4.2: Submerge the Conductivity sensor into the
calibration solution to the second notch.
The sensor must be centered in the cal-
ibration beaker and is not permitted to con-
tact the bottom of the calibration beaker.
5.
5.1: Compare the
Nom. val.
(circle) with the
value on the calibration solution.
Pressing the
Nom. val.
button (circle)
causes a numerical entry field to appear
where the nominal value can be adjusted.
5.2: Wait until the temperature value (circle) is
stable.
Recalibration is performed only when
the temperature value is stable.
The nominal value is factory set at 2%/°C
temperature compensation (based on 25 °C).
(
Meas.Channels Conduct\Temp.Comp
menu.)
If the temperature compensation is switched
off, the actual value at the actual tempera-
ture must be compared to the value on the
calibration solution table.