Instruction Manual AquaMaster
Servicing
94
12745E/4
8.2.5
Automatic adjustment with AquaScat 2 WTM A
The automatic adjustment is possible only with instrument type AquaScat 2 WTM A. The
automatic adjustment can also be set for time actuation in the
Adjust interval
menu. This
is described in the Reference Manual.
The following procedure describes how manual triggering of an adjustment is made with an
AquaScat 2 WTM / HT:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Switch the photometer to service operation.
Section 6.10
2.
Press the
Local ……
button.
3.
Press the
Recalibration
button and then
C1
Turb
.
4.
Carry out the adjustment as follows:
Press the
initiate...
button. This starts the ad-
justment.
If the adjustment was successful, this is confir-
med with
Adjustment OK
. This completes the
adjustment.
If the calibration check was
not
successful, it is
indicated with
Adjust. fault
. The following
points can now be checked:
Check the cleanliness of the optics accord-
ing to Section 8.2.7.
Check whether the outlet cone is correctly
mounted according to Section 8.2.6.
After performing the checks above, repeat the
adjustment.
If the adjustment could not be success-
fully completed, contact your country represen-
tative.
A new recalibration factor is determined during the adjustment. The deviation from the
original state is displayed under
Curr. corr.
.