Instruction Manual AquaMaster
Settings
60
12745E/4
7.6
Set outputs
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Local …….
button.
4.
Press the
Inp./outputs
button.
If the desired menu does not
appear, press the arrow key bottom
right.
5.
Press the
Outputs
button.
6.
Select
O1 ..
O8
.
7.
Activate outputs (multiple selection possible).
Activated outputs are highlighted
green.
Invert: inverts the outputs
Prio fault
Fault
Warning
Service
Adjustment
Limit 1 .. 8
The other buttons named
MR-
Out…
and
Valve/Channel
are for
automatic measuring range switch-
ing and for multiple sample switch-
ing with valves.
Reference Ma-
nual.
8.
Press the
Meas
button.
Instrument again in measuring op-
eration.