Servicing
Instruction Manual AquaMaster
12745E/4
79
WORKSTEP
ADDITIONAL INFO / IMAGES
5.3: Press the
Initiate
button. The recalibration
begins.
If the adjustment was successful, confirm with
Adjustment OK
. This completes the adjust-
ment.
If the adjustment was not OK, the following
messages may appear:
running...
Cause: Values not yet stable.
Out of tolerance
Cause: Current actual value is too far from the
nominal value.
Action:
Ensure that the set nominal value and the
nominal value of the calibration solution
match.
Clean sensor.
If the quality indication af-
ter calibration is between 100
and 35, the age of the sensor is
the reason.
If the calibration was not correct,
30 is shown at quality. The cali-
bration must be repeated.
6.
Rinse sensor with distilled water.
7.
Install sensor in measuring cell block in accord-
ance with Section 8.1.3.