Servicing
Instruction Manual AquaMaster
12745E/4
77
8.1.6
Clean and calibrate Redox/ORP sensor
CAUTION!
The Redox/ORP sensor can be damaged through improper handling.
The Redox/ORP sensor can be damaged by carelessly touching the electrode or by using the
wrong cleaning agent.
For cleaning this sensor please consult Section 8.1.1.
Touch the measuring tip of the Redox/ORP sensor only if absolutely necessary.
Do not clean the sensor with abrasive cleaning agents.
Do not clean the sensor when dry.
Use only recommended cleaning agents for cleaning.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
1.1: Press the
Menu
button.
1.2: Set access code and confirm with
OK
.
Factory setting is
0
.
1.3: Press button with
(S1 .. S4) name of the
Redox/ORP sensor
.
1.4: Select the
Recalibration
menu.
1.5: Select
C1 ORP
menu.
2.
Remove Redox/ORP sensor in accordance with
Section 8.1.2 and position in the locking de-
vice.
3.
Clean measuring tip of the sensor.
3.1: Dip the measuring tip into cleaning solu-
tion or dab with a soaked cloth.
Use cleaning agents in accordance with Section
8.1.1.3.
3.2: Rinse the measuring tip with distilled wa-
ter and dab dry.
4.
Prepare recalibration.
4.1: Open calibration holder and fill with buffer
solution by pressing the container.
Calibration solutions from various
manufacturers are supported. They can be
selected in the
Recalibr./Cali. standard
menu. The one from Hamilton is set as stan-
dard.