Servicing
Instruction Manual AquaMaster
12745E/4
101
WORKSTEP
ADDITIONAL INFO / IMAGES
8.
Restore the service voltage to the photome-
ter.
On instruments
without
automatic adjust-
ment, carry out a manual adjustment as de-
scribed in Section 8.2.4. This servicing duty is
then completed.
On instruments
with
automatic adjustment
(AquaScat 2 WTM A) continue with step 9.
9.
Change over the instrument to service opera-
tion as described in Section 6.10.
10. In the
Local ……/Simulation
menu, select
the
Motor position
submenu. Now select
the
Adjustment
menu item.
The checking unit now moves to the adjust-
ment position.
11. Remove the optics unit of the photometer
and fasten onto the docking station.
Section 8.2.1
12. Clean the head of the checking unit.
12.1: Remove the two screws (arrows) and
then remove the head of the checking unit.
12.2: Clean the two glass panels (arrows)
with a cotton-tipped applicator moistened
with ethanol.
Position glass
panel 1
Position glass
panel 2