Instruction Manual AquaMaster
Operation
52
12745E/4
6.10
Switching to service operation
The system is configured in service operation. The measuring procedure is interrupted and
the main menus appear on the display. Proceed as follows to access service operation:
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set access code and confirm with
OK
.
Factory setting is
0
.
3.
Select menu
Local …….
or
S 1 .. 8
.
Now the instrument is in service op-
eration.
The following applies in service operation:
* The measuring values remain on the last values on the digital interfaces.
* Depending on the configuration, the current outputs go to 0/4 mA or remain on the
last measuring value.
The limits are deactivated.
If an output for service is programmed, it is activated.
Error messages are suppressed.
* This applies when the
Local parameters\Current outputs\General\For service
is set to
Measure
.
For measuring operation press the
Meas
button. When switching from service operation to
measuring operation, an hourglass appears in the information bar for about 20 seconds.
The measuring values are frozen during this time.