Instruction Manual AquaMaster
Troubleshooting
110
12745E/4
The following error messages can be displayed:
ERROR MESSAGE
DESCRIPTION
POSSIBLE CAUSES
V ANALOG
One of the internal analogue
voltages is outside the permit-
ted range.
Defect in the electronic system.
Service technician
MEASUR.FAULT
Measurement value acquisi-
tion is faulty.
Air bubbles in the water.
External light in the vicinity of
the measuring station (e.g.
transparent hoses).
Defect in the electronic system.
Service technician
AN.MEAS.FAULT
The measurement value ac-
quisition of the analogue
channels is disturbed.
Defect in the electronic system.
Service technician
LIGHT SOURCE 1
The detector for monitoring
the LED receives no light.
Defective light source.
Service technician
ANALOG IN 1/2
The input signal on analog in-
put 1/2 is less than the error
limit.
There is no input signal.
POWER LINK
Actuation of the extended in-
puts/outputs via the Powerlink
is disturbed.
Interrupted connection to the
extended inputs/outputs.
MEASUR.FAULT
Serious measuring fault on a
Hamilton sensor.
Sensor measurement defective.
Temperature measurement de-
fective.
Resistances or electric poten-
tials outside the permitted
range.
CALIBRATION
Serious fault calibrating a
Hamilton sensor.
Oxygen sensor: cap missing.
pH, Redox/ORP sensor: sensor
defective (quality <15%).
Conductivity sensor: sensor de-
fective (quality <15%) or out-
side medium.
INTERFACE
Connection problem with
Hamilton sensor.
Oxygen sensor: current output
fault.
HARDWARE
Serious fault in the hardware
of a Hamilton sensor.
Input voltage far outside the
permitted range.
Temperature measurement far
outside the permitted range.
Oxygen sensor: red channel
failure.
Internal communication fault.
Table 5: Possible fault messages