Instruction Manual AquaMaster
Servicing
92
12745E/4
8.2.4
Manual adjustment
The following procedure describes how manual adjustment is made with an AquaScat 2
WTM / HT:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the sample supply to the photome-
ter.
2.
Loosen the conduit gland (X) and remove the
inlet pipe (Y).
3.
Remove the optics unit of the photometer
and fasten onto the docking station.
Section 8.2.1
4.
Inspect the checking unit to ensure it is clean.
5.
Mount the checking unit and fasten with the
knurled screws.
Make sure that the pins enter the positioning
holes (circles).