Decommissioning/storage
Instruction Manual AquaMaster
12745E/4
113
11
Decommissioning/storage
11.1
Decommissioning
The aim of decommissioning is to prepare the individual components of the system properly
for storage.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage to the system.
2.
Interrupt main water supply.
3.
Remove front cover of the AquaScat 2and re-
move electrical connections.
Section 4.7
4.
Re-fasten front cover onto the AquaScat 2.
5.
Remove the optics unit of the AquaScat 2 and
fasten onto the docking station.
Section 8.2.1
6.
Remove junction hoses and then clean and dry
the measuring cell housing of the AquaScat 2.
7.
Re-mount optics unit on the AquaScat 2.
8.
Remove the AquaScat 2 from the base plate
and close all openings.
9.
Unscrew sensor connection cables to the con-
nection box and then remove connection box
from the contact surface of the base plate and
pack it.
10. Remove sensors from the measuring cell block,
clean them, and then pack according to the
manufacturer's instructions.
Store the measuring tips of the pH and
Redox/ORP sensors in containers with clean tap
water so that they are protected against drying
out.
The protection covers intended for that pur-
pose with the 3 molar solution of potassium
chloride can also be used.
Section 8.1.1.3
11
Disassemble and clean tubing with optional
components (e.g. deaeration tube, flow me-
ter).
Section 8.1.11
12. Remove the base plate from the wall and pack
for storage.