Instruction Manual AquaMaster
Servicing
80
12745E/4
8.1.7
Cleaning and calibrating the oxygen sensor
CAUTION!
The oxygen sensor can be damaged through improper handling.
The oxygen sensor can be damaged by touching the electrode carelessly or by using incor-
rect cleaning agents.
See Section 8.1.1 for details on how to clean the sensor.
Only touch the measuring tip of the oxygen sensor when absolutely necessary.
Do not clean the sensor with abrasive cleaning agents.
Cleaning should not be carried out without cleaning agent.
Only use the recommended cleaning agents.
Before calibrating in air, the current ambient air pressure must be known (barometer or in-
formation from weather service –
NOTE
: use the QFE value).
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
1.1: Press the
Menu
button.
1.2: Set the access code and confirm with
OK
.
Factory setting is
0
.
1.3: Press the button with
(S1 .. S4) Name of
the oxygen sensor
.
1.4: Select the
Recalibration
menu.
1.5: Select the
C1 Oxygen
menu.
2.
Remove the oxygen sensor according to Sec-
tion 8.1.2 and position it in the lock.
3.
Clean the measuring tip of the sensor.
3.1: Immerse the measuring tip in cleaning so-
lution or dab it with a soaked cloth.
Use the cleaning agent according to Section
8.1.1.3.
3.2: Rinse off the measuring tip with distilled
water and pat dry.