Instruction Manual AquaMaster
Servicing
82
12745E/4
8.1.9
Installing an unconfigured sensor
This process is only applicable if a new sensor has not been ordered from SIGRIST-
PHOTOMETER.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
In the
Local ……
menu, access the
Digi. interf.
submenu.
2.
Remove the old sensor from the measuring cell
block according to Section 8.1.2.
3.
Remove the holder from the old sensor and
screw it onto the new sensor.
On pH and redox/ORP sensors, the cap with
the storage solution must be removed.
4.
Install the new sensor in the measuring cell
block according to Section 8.1.3 and attach
the connection cable.
5.
In order for the assignment of the slave num-
ber to be carried out reliably, detach the
connection cables of all other sensors.
6.
Select the
Hamilton
menu.
Press
start…
under the “Find Sensor” menu
item.
7.
A system search is now carried out for connec-
ted Hamilton sensors. As soon as a sensor is
found, the type and slave number are display-
ed (e.g. Oxygen, Slave no. 1).
If the slave number has not yet been adjusted
to Siginet, then
undefined
is shown under the
slave number menu item.
In this case, a slave number has to be entered
according to the following table.
Oxygen
pH
Conduc-
tivity
Re-
dox/ORP
1
2
3
4
8.
The other sensors can now be reattached.