Application examples
C.2 Example 1: Liquid resin in storage vessel, level measurement
SITRANS Probe LR with mA/HART
Operating Instructions, 05/2019, A5E32337711-AC
95
①
Reference point
②
Span
③
100% Level
④
0% Level
⑤
Empty
Parameter
Description
Value
Function
P001
Mode of measurement
1
Level
P003
Measurement response
2
1 m/minute
P005
Units
1
meters
P006
Empty distance
5
5 m
P007
Span
4.5
4.5 m
P070
Failsafe timer
2
2 minutes
P071
Failsafe mode
1
Hi
P838
Auto False-Echo Suppression
Distance
Distance
1)
minus 0.5 m Sets length of learned TVT
curve
2)
to use
P837
Auto False-Echo Suppression 2, then 1
Enables the use of learned
TVT curve
2)
1)
Distance to product from reference point.
2)
For more details on setting P837 and P838, see P837 Auto false echo suppression
(Page 66). For an explanation, see Auto false-echo suppression (Page 88).
Return to RUN: press PROGRAM
to start normal operation.