Parameter assignment
7.27 TVT (Time Varying Threshold) adjustment parameters (P831 to P839)
SITRANS Probe LR with mA/HART
66
Operating Instructions, 05/2019, A5E32337711-AC
7.27
TVT (Time Varying Threshold) adjustment parameters (P831 to
P839)
Note
Factory settings are indicated by an asterisk (*) in the parameter tables, unless explicitly
described.
First SITRANS Probe LR learns the echo profile. Then the learned profile, or part of the
learned profile, is used to screen out false echoes.
The following parameters are for authorized Siemens Service personnel or technicians
familiar with Siemens echo processing techniques. View the echo profile first, before
attempting to modify these parameters.
7.27.1
P831 Manual TVT Shaper
Note
This parameter is for use only by Siemens service personnel.
Turns the TVT Shaper ON or OFF.
7.27.2
P832 Manual TVT Shaper Adjust
Note
This parameter is for use only by Siemens service personnel.
7.27.3
P837 Auto false echo suppression
Note
•
P837 and P838 should be used together to adjust the TVT (Time Varying Threshold)
curve.
•
Use this function only if there is a minimum distance of 2 meters (6.5 feet) from the
reference point to the material.
•
Set P837 and P838 during start up, if possible.
•
If the vessel contains an agitator, the agitator should be running.