Commissioning
6.2 Local commissioning
SITRANS Probe LR with mA/HART
Operating Instructions, 05/2019, A5E32337711-AC
37
①
Reference point
②
Span
③
Level setup
④
Level
⑤
Empty
Note
P006 and P007 are interlinked: see notes attached to P007.
6.2.11.6
P007: Span
Set the range to be measured (P007: Span)
Value
Range
0.0000 to 20.00
Default
19.56 (see note below)
Enter the distance between Empty (process empty level) and Span (process full level), in the
units set in P005. Span can be set at any distance above Empty level.
Note
•
Setting P006 resets Span, if it has not previously been set to a different value.
•
The default setting for Span is based on Operation (P001) and Empty (P006). Span is set
to Empty minus 110% of Blanking distance, unless Operation is set to distance (P001=3).
In this case, Span is set to Empty distance. Blanking distance is 0.3 m (1 ft), plus the
shield length (if any).
•
Always prevent the monitored surface from approaching within 0.3 m (1 ft) of the
reference point, as this is the minimum distance detectable.