Parameter assignment
7.20 Range calibration (P652 to P655)
SITRANS Probe LR with mA/HART
Operating Instructions, 05/2019, A5E32337711-AC
57
7.19.2
P342 Power-on resets
The number of times power has been applied since the date of manufacture.
Values (view only)
Range
0 to 99999 (days)
Related parameters
P341 RUN Time
This parameter is updated every time the instrument is reset or is powered up.
7.19.3
P343 Internal temperature
WARNING
Internal temperature
Internal temperature must not exceed 80 °C (176 °F).
Displays (in degrees C) either the current temperature on the circuit board, or the maximum
or minimum temperature recorded by the internal sensor. The high and low values are
maintained over a power cycle.
Values (view only)
Range
–50 °C to 150 °C
1
Current temperature
2
Maximum temperature
3
Minimum temperature
7.19.4
P346 Serial number
Displays the serial number of the instrument. The numbers stored in Index 2, followed by the
numbers stored in Index 1, give you the complete serial number.
Index 2
Index 1
Values (view only)
Range: 00000 to 99999
Range: 00000 to 99999
Example: 1503010
15
03
010
7.20
Range calibration (P652 to P655)
Note
Factory settings are indicated by an asterisk (*) in the parameter tables, unless explicitly
described.