EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Clearing Malfunctions
15.1
If the EuroFID does not work at all …
15.2
If the measured values are obviously incorrect …
Possible cause
Note
The mains cable is not correctly
connected.
Check the mains cable and the connections.
The main switch is set to OFF.
Check the mains switch of the EuroFID (rear of
the operating unit).
Mains supply failure.
Check the external switches and fuses.
Check the mains voltage.
The internal mains fuse is
defective.
Check the internal mains fuse (
).
Internal operating temperatures
are incorrect.
Check the malfunction messages (
The supply of sample gas is not
working.
Check the supply of sample gas (
The internal software is not
working.
Switch off the EuroFID operating unit and switch
on again after a few seconds.
The overtemperature switch of the
analyzer unit has been triggered.
Possible cause
Note
The EuroFID is not ready for
operation.
Follow the start-up procedure (
Pay attention to the status messages (
).
The EuroFID is not correctly
calibrated.
Check the employed span gas (
).
Check the setpoint value (
If necessary, re-run calibration.
The averaging setting is too high.
Check (
); if necessary change
for testing purposes.
The pressure at the sample gas
inlet is either too high or too low.
Check the supply of sample gas (
The sample gas supply is leaking.
Inspect.
Run a leakage test on difficult installations.
If only observed on an analog
output:
The load is too high.
Ensure that the internal resistance of the
connected devices is not greater than 500
.
The EuroFID measuring system is
soiled.
Contact the manufacturer's after-sales service
department.
The stability of the zero and calibration point depends on the consistency of
the temperature of the gas-carrying components.
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...