176
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Clearing Malfunctions
15.5.2
Possible malfunction messages
N Spontaneous message
Message text in the error archive
Significance
Duration of the
message
Countermeasures
1
P r o g r . E r r o r
S Y S T E M E R R O R
A p r o g r a m b u g h a s
o c c u r r e d .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
EuroFID Switch off and on
once more.
If unsuccessful:
Have the
EuroFID serviced by the
after-sales service of the
manufacturer.
2
S o f t w a r e b u g
S Y S T E M E R R O R
A p r o g r a m b u g h a s
o c c u r r e d .
3
D i v . b y _ 0
S Y S T E M E R R O R
D u r i n g c a l c u l a t i o n a
d i v i s i o n b y z e r o h a s
o c c u r r e d .
4
A r i t h m O v e r f l
S Y S T E M E R R O R
D u r i n g a c a l c u l a t i o n
a n o v e r f l o w h a s
o c c u r r e d .
5
W a t c h d o g e r r
S Y S T E M E R R O R
T h e w a t c h d o g c o n t r o l
h a s t r i g g e r e d a
r e s e t .
A reset has been
automatically activated
after a program error in
the firmware.
Measurement is
temporarily interrupted.
The short message
text will normally not
be visible as the
program restarts
again immediately.
Wait until the EuroFID is
once more ready for
operation.
6
P a r a m e t e r e r r
S Y S T E M E R R O R
T h e p a r a m e t e r s a r e
l o s t . R e c o n f i g u r a t e
t h e s y s t e m .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
7
P a r a m _ e r r o r
S Y S T E M E R R O R
A n a l y z e r p a r a m e t e r s
a r e w r o n g o r l o s t .
R e c o n f i g u r a t e t h e
p a r a m e t e r s o f t h e
a n a l y z e r
8
G P a r a m . e r r o r
S Y S T E M E R R O R
T h e l a s t c h a n g e i n
p a r a m e t e r s i s l o s t .
P l e a s e c o n f i g u r a t e
a n d r e l e a s e t h e
p a r a m e t e r s
9
E P R O M T y p e
S Y S T E M E R R O R
T h e F l a s h - E p r o m t y p e
i s w r o n g . P l e a s e
c h a n g e t o t h e
c o r r e c t F l a s h - E p r o m
t y p e
Internal fault.
Measurement is not
possible.
Can only occur
during
manufacturing.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
10
E P R O M N O
S Y S T E M E R R O R
T h e e n t e r e d n u m b e r
o f F l a s h - E p r o m i s
w r o n g . P l e a s e c h a n g e
a n d r e l e a s e t h e
s y s t e m p a r a m e t e r s .
11
E P R O M B L K 1 E r
S Y S T E M E R R O R
T h e p a r a m e t e r s o f
b l o c k 1 a r e n o t u p
t o d a t e . T h e m e m o r y
i s p o s s i b l y
d e f e c t i v e . T h e c o p y
o f t h e p a r a m e t e r s i n
b l o c k 2 i s O K .
Internal malfunction
which is compensated
by internal backup
functionality.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
12
E P R O M B L K 2 E r
S Y S T E M E R R O R
T h e p a r a m e t e r s o f
b l o c k 2 a r e n o t u p
t o d a t e . T h e m e m o r y
i s p o s s i b l y
d e f e c t i v e . T h e c o p y
o f t h e p a r a m e t e r s i n
b l o c k 1 i s O K .
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...