38
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Analyzer Unit
4.1
Installation of the “wall housing” type analyzer unit
4.1.1
Assembly/securing
Mount the “wall housing” analyzer unit on a stable basis (any installation position).
Ensure that the installed sample gas filter remains accessible (
If a heated sample gas line is used:
Take the space requirements of the sample gas
line into account.
On devices with the “flow barrier” option:
Secure the supplied solenoid valve at a
suitable location.
4.1.2
Gas connections
Gas connection
for
Installation work
Instr. air
Instrument
air
Observe the information on instrument air (
).
Connect the instrument air supply to the “Instr.air”
gas connection.
On devices with the “flow barrier” option:
Also connect the instrument air supply to a gas
connection of the solenoid valve.
Connect the other gas connection of the solenoid
valve to the “flow barrier” gas connection.
Fuel gas
Fuel gas
Observe the information on fuel gas (
Connect the fuel gas supply to the “Fuel gas” gas
connection.
Calibr. gas
Span gas
Provide suitable span gas (
).
Observe the information on span gas supply (
).
Connect the span gas supply to the “Calibration gas”
gas connection.
Clamping ring
screw
connection on
the protruding
block
Sample gas
inlet
Observe the notes on sample gas supply (
).
Connect the sample gas inlet to the sample gas
sampling point. Suitable material: Non-corrosive
metal pipe with an outer diameter of 6 mm.
Cylindrical hose
connection on
the protruding
block
Bypass
If it is necessary to eliminate the bypass flow due to the
composition of the sample gas:
Connect an exhaust gas line to the “Bypass”
connection. Suitable material: Non-corrosive hose
with an inside diameter of 16 mm.
Observe the information on installation of the
exhaust gas connections (
).
Exhaust
Exhaust gas
If it is necessary to eliminate the exhaust gas due to the
composition of the sample gas:
Connect an exhaust gas line to the “Exhaust”
connection. Suitable material: Non-corrosive hose
with an inside diameter of 6 mm.
Observe the information on installation of the
exhaust gas connections (
).
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...