112
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
“Maintenance” Sub-menu Functions
11.6
Filter change (menu)
Function
The “Filter change” menu function assists you during maintenance work on the sample gas
filter in two steps:
a) Safeguarding before starting maintenance:
– The supply of instrument air to the ejector pump is stopped. As a result the ejector
pump is deactivated; the supply of sample gas to the measuring system is stopped.
– The supply of fuel gas is stopped, the FID flame is extinguished.
–
On devices with the “flow barrier” option:
The external solenoid valve is activated;
as a result the sample gas in the sample gas filter is displaced by instrument air.
b) Safe restart:
– The sample gas filter is heated to operating temperature; this prevents
condensation.
– The FID flame ignites again.
–
On devices with the “flow barrier” option:
The external solenoid valve is deactivated.
– The sample gas supply is activated again.
Open the menu
Main menu
Maintenance
Filter change
Procedure (abbreviated)
a)
Before commencing maintenance work:
Open the “Filter change” menu (Main menu
Maintenance
Filter change).
Set the “Filter change” status to “YES” (Selection box 1).
Close the “Filter change” menu (”END”, enter key).
b)
After maintenance work:
Open the “Filter change” menu once more.
Set the “Filter is replaced” status to “YES” (Selection box 2).
Close the “Filter change” menu (”END”, enter key).
Wait until measurement mode is automatically activated once more (approx.
15 min.).
For a detailed description the operating steps and maintenance work refer to
M a i n t e n a n c e
F i l t e r c h a n g e
F i l t e r c h a n g e
N O
F i l t e r i s r e p l a c e d
N O
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
S y s t e m i n S e r v i c e
Menu title
Selection box 1
Selection box 2
Current measured value
Current measuring range (measuring range end value)
E n d
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...