“Maintenance” Sub-menu Functions
Settings for automatic measuring range switch-over
1
Press
or
until the selection box is highlighted.
2
Press
until
a u t o
or
a u t o A O u
is displayed (functional difference
).
3
If necessary:
Modify the settings.
a) Press
/
until the respective entry field is highlighted.
b) Enter the required value.
To finish:
4
Press
or
until
E n d
is highlighted.
5
Press .
R a n g e S e l e c t M o d e
M e a s u r i n g r a n g e s w .
R a n g e : M i n i m u m = 1
R a n g e : M a x i m u m = 6
M a x i m u m M i n i m u m
9 9 % 9 %
D e l a y
0 . . . 9 0 . 0 1 . 0 s
C h a n g e o r E n d
F a c t o r y l e v e l a c t i v e
R a n g e . . . 1 0 0
S y s t e m i n S e r v i c e
Menu title
Selection box (selectable modes
)
Entry field 1
Entry field 2
Entry field 3, entry field 4
Entry field 5
Current measured value
Setting
Function
R a n g e : M i n i m u m
The smallest measuring range with automatic measuring range
switch-over
R a n g e : M a x i m u m
The largest measuring range with automatic measuring range switch-
over
M a x i m u m
Threshold value at which the EuroFID switches to the next largest
measuring range
1
1
Relative to the measuring range end value of the current measuring range; the effect is only valid after the set delay
M i n i m u m
Threshold value at which the EuroFID switches to the next smallest
measuring range
D e l a y
Delay time for automatic measuring range switch-over: The
measuring range is only switched over if the output value for the
duration of the delay time is outside the threshold value.
2
2
Setting range: 0.0 … 99.9 seconds. The maximum value is, however, 90 s, even if a higher value is set here.
Set “Minimum” and “Maximum” so that a switching hysteresis occurs.
Otherwise the EuroFID would constantly switch between the measuring ranges
if the measured value corresponds to the threshold value.
●
It depends on the application which settings are the most suitable.
●
Standard value for the switching hysteresis: 2 %.
●
Make the switching hysteresis greater if you expect fluctuating or “noisy”
measured values.
a u t o
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...