Installation of the Analyzer Unit
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
43
4.3.2
Assembly
Installation site
●
Comply with the permissible ambient temperature (specification
).
●
The installation site must be free from strong vibrations and jolts (specification
●
The sampling probe must protrude freely into the process gas flow.
Fig. 8
Exhaust gas discharge on the sampling probe of the analyzer unit type “inline”
Preparing for assembly
At the installation site install a mounting connection (counterpart to the flange of the
analyzer unit). Ensure that the mounting connection can bear the weight of the analyzer
unit. If necessary support the mounting connection.
Securing
1
Position a flange seal between the mounting connection and the flange of the analyzer
unit.
If the analyzer unit is used for hot processes (> 300 °C):
Use the heat insulation
flange instead of the flange seal (
If the hood for weather protection is to be fitted:
Fit the installation plate of the
hood for weather protection between the mounting connection and flange of the
analyzer unit (
2
Carefully insert the probe of the analyzer unit through the flange pipe.
3
Connect the flanges with the supplied screw fittings.
NOTICE:
The bypass exhaust gas flow is discharged from the surrounding tube of the
sampling probe.
Position the sampling probe so that the exhaust gas can flow freely and is
not drawn in by the sampling probe.
Never install any components so near to the sampling probe that the
EuroFID exhaust gas could be drawn in by the sampling probe.
Otherwise, incorrect values could be measured.
●
Mounting connection (accessory)
●
Weight (mass) of the analyzer unit type “inline”
●
Additional weight of the hood for weather protection (accessory): approx.
13 kg
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...