Technical Data
16.3
Ambient conditions
Analyzer unit (except “industrial housing” type)
Ambient temperature:
+5 … +45 °C (41 … 113°F)
Storage temperature:
–20 … +70 °C (–4 … 158 °F)
Relative humidity:
80 % max. brief superficial moisture condensation permissible
Ambient air pressure:
700 … 1200 hPa
Atmospheric influences:
●
Install indoors.
●
Prevent exposure to direct sunlight.
●
Protect from aggressive atmosphere.
Vibration/jolts:
●
Installation site free from pronounced vibration and jolts
●
Acceleration < 0.2 m/s
2
at f =10 … 150Hz
Degree of soiling:
1
Analyzer unit type “industrial housing”
Ambient temperature
Standard version:
–30 … +30 °C (–22 … +86 °F)
With air conditioner (option):
–30 … +50 °C (–22 … +122 °F)
Storage temperature:
–20 … +70 °C (–4 … 158 °F)
Relative humidity:
5 … 90 %, external moisture condensation permissible
Ambient air pressure:
700 … 1200 hPa
Atmospheric influences:
●
Installation site partially protected against weather.
●
Exposure to direct sunlight is not permissible.
●
External moisture condensation permissible.
Vibration/jolts:
The installation area must be free from vibrations and jolts.
Degree of soiling:
> 1
Operating unit
Ambient temperature:
+0 … +40 °C (32 … 104 °F)
Storage temperature:
–20 … +60 °C (–4 … 140 °F)
Relative humidity:
5 … 90 %, external moisture condensation not permissible
Atmospheric influences:
●
Use indoors.
●
Completely protected against the weather.
●
Protect from aggressive atmosphere.
Vibration/jolts:
The installation site must be virtually free from vibration.
Degree of soiling:
1
Terminal box
Ambient temperature:
–10 … +55 °C (14 … 131 °F)
Storage temperature:
–25 … +70 °C (–13 … +158 °F)
Relative humidity:
5 … 90 %, external moisture condensation permissible
Atmospheric influences:
●
Installation site partially protected against weather.
●
Protect from aggressive atmosphere.
Vibration/jolts:
The installation site must be virtually free from vibration.
Degree of soiling:
> 1
●
Weather-protected installation as stipulated in EN 60721-3-3
●
Degree of soiling 1 = no soiling or only dry, non-conductive soiling
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...