“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
117
11.10
Temperature of the sensor block
Function
The sensor block is heated to prevent condensation in the parts in contact with the sample
gas. The set temperature of the heater is adjustable. It depends on the measuring
application which set temperature is suitable. Standard value: 200 °C.
For applications in which the sample gas is taken from potentially explosive areas, the
setpoint temperature must be set to a value which is at least 25 °C below the
classification temperature (refer to type plate).
Example:
Classification temperature: 220 °C; sensor block set temperature:
195 °C
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
Setup
temp. sensbl.
2
Press
or .
➝
Entry field is highlighted.
3
Enter the required value.
4
Press
or .
➝
E n d
is highlighted.
5
Press .
Do not change the setpoint temperature set at the works.
S y s t e m P a r a m e t e r s
S e t u p t e m p . s e n s b l .
S e n s o r b l o c k s e t t e m p
1 9 0 . 0 0 C E L
C E L
u p : 6 0 . 0 0
u p t o : 2 3 0 . 0 0
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Entry field
Minimum value
Maximum value
Current measured value
Current measuring range (measuring range end value)
E n d
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...